Event Handling Principles
Introduction to the System
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
45
Event Handling Principles
The HP 81200 system can detect a variety of events and react on events.
The reaction may simply be a trigger pulse at the TRIGGER OUTPUT of the
clock module, but can also be a change of the test sequence.
“What is an Event?” on page 46
“Actions Upon an Event” on page 47
Usage of Events
Reacting on events provides many capabilities:
•
Stop and go:
This is useful for production tests, where data is sourced to the DUT, a
measurement is performed with other equipment, the next data pattern
is sourced, and so on.
•
Block switching:
The data sequence is no longer fixed. Based on certain events, certain
portions of the overall sequence can be executed.
This has the advantage that one and the same sequence can be created
and downloaded once and then used for several tests. There is no need
for re-programming the instrument.
•
Trigger external devices:
The event can generate a trigger signal at the TRIGGER OUTPUT of the
clock module. This can be used to trigger an external instrument like a
sampling oscilloscope or logic analyzer to sample the data at an error
location.
•
Bolt on:
The HP 81200 system can be integrated into a large IC test system. The
IC tester would issue a trigger to start the HP 81200 system for a special
measurement. The HP 81200 system would perform the test and return
pass/fail information that can be examinated and evaluated by the IC
tester.
•
Match loop:
PLL-based devices typically require an initialization segment that has to
be repeated until the device is synchronized. The event that controls
repetition would be “an error occurred”.
Содержание 81200
Страница 1: ...Revision 2 1 HP 8 1 200 D a t a G e n e r a t o r A n a l y z e r P l a t f o r m User Guide ...
Страница 9: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 9 What s New in This Release ...
Страница 15: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 15 Introduction to the System ...
Страница 49: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 49 Test Development Overview ...
Страница 57: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 57 Structure of the User Interface ...
Страница 83: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 83 Setting Global System Parameters ...
Страница 93: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 93 Connecting the DUT ...
Страница 105: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 105 Setting Up Ports and Channels ...
Страница 127: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 127 Choosing the Kind of Measurement ...
Страница 155: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 155 Creating and Editing Segments ...
Страница 171: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 171 Using the Data Sequence Editor ...
Страница 181: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 181 Running the Test ...
Страница 186: ...Running the Test How to Start Stop the Test 186 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...
Страница 187: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 187 Viewing Generated and Captured Data ...
Страница 199: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 199 Using Auxiliary Functions ...
Страница 215: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 215 How Do I ...
Страница 240: ...Index 240 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...