How to Compensate for Internal and External Delays
Using Auxiliary Functions
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
203
3
Click the Measure button.
For generators, the following window pops up:
Figure 114
Deskew Editor—Zero Adjust Window for a Generator Channel
4
Follow the instructions given in this window.
Enter the master cable delay in the provided field. The typical delay of a
1 m (3 feet) SMA cable is about 3.5 ns.
The zero adjustment of analyzers requires a signal generator to be used
as a reference channel. The system proposes the first of the installed
generator channels. If there is no generator installed, the TRIGGER
OUTPUT of the clock module is suggested.
The measured zero adjust value is automatically entered in the table and
the next instruction to connect the next generator output is displayed.
When all marked generators are adjusted the Automated Deskew window
disappears.
How to Compensate for Cable Delays
The cable delay compensation assures that the edges of all generator
outputs of the HP 81200 system appear at the same time at the end of the
cables used in the setup. For the HP 81200 system analyzer inputs the cable
delay compensation assures that all output signals of the DUT are sampled
at the same time at the end of the cables, close to the DUT.
N O TE
To perform this procedure, you need all the cables which are going to be
used.
The cable delay compensation range is ±23 ns.
Perform the cable delay compensation first for the generator outputs, then
for the analyzer inputs.
It is recommended to use cables of equal type and length for all inputs and
outputs.
Содержание 81200
Страница 1: ...Revision 2 1 HP 8 1 200 D a t a G e n e r a t o r A n a l y z e r P l a t f o r m User Guide ...
Страница 9: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 9 What s New in This Release ...
Страница 15: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 15 Introduction to the System ...
Страница 49: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 49 Test Development Overview ...
Страница 57: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 57 Structure of the User Interface ...
Страница 83: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 83 Setting Global System Parameters ...
Страница 93: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 93 Connecting the DUT ...
Страница 105: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 105 Setting Up Ports and Channels ...
Страница 127: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 127 Choosing the Kind of Measurement ...
Страница 155: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 155 Creating and Editing Segments ...
Страница 171: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 171 Using the Data Sequence Editor ...
Страница 181: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 181 Running the Test ...
Страница 186: ...Running the Test How to Start Stop the Test 186 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...
Страница 187: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 187 Viewing Generated and Captured Data ...
Страница 199: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 199 Using Auxiliary Functions ...
Страница 215: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 215 How Do I ...
Страница 240: ...Index 240 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...