Using Auxiliary Functions
How to Compensate for Internal and External Delays
206
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
6
Activate
Probe
.
The following window pops up:
Figure 116
Deskew Editor—Probe Measurement Window for a Generator Channel
7
Enter the
Probe Attenuation
factor
For the HP 1144A Active Probe this is 10.
8
Enter the
Probe Delay
.
If you don’t know the probe delay, you can make a first measurement by
connecting the probe directly to the first generator output. Start the
measurement by clicking OK. The result is the probe’s propagation
delay.
Close the Automated Deskew window. Click
Measure
again. Enter the
resulting value as the probe delay and repeat the measurement. Now the
resulting value for this output should be 0 ns.
9
Follow the instructions given in the window.
The measured delay value is automatically entered into the table and the
next instruction to connect the next generator output is displayed. After
all marked generators have been adjusted the Automated Deskew
window disappears.
10
Continue with the analyzers.
This is essentially the same procedure, except that you don’t use the
PROBE connector of the clock module, but one of the generator
channels as a reference. The system proposes the first of the installed
generator channels.
N O T E
If you are using the high speed differential analyzer HP E4837A, please
note:
The analyzer must be operated in differential mode.
Both inputs must be connected to a differential generator (E4838A or
E4843A).
Содержание 81200
Страница 1: ...Revision 2 1 HP 8 1 200 D a t a G e n e r a t o r A n a l y z e r P l a t f o r m User Guide ...
Страница 9: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 9 What s New in This Release ...
Страница 15: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 15 Introduction to the System ...
Страница 49: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 49 Test Development Overview ...
Страница 57: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 57 Structure of the User Interface ...
Страница 83: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 83 Setting Global System Parameters ...
Страница 93: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 93 Connecting the DUT ...
Страница 105: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 105 Setting Up Ports and Channels ...
Страница 127: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 127 Choosing the Kind of Measurement ...
Страница 155: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 155 Creating and Editing Segments ...
Страница 171: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 171 Using the Data Sequence Editor ...
Страница 181: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 181 Running the Test ...
Страница 186: ...Running the Test How to Start Stop the Test 186 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...
Страница 187: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 187 Viewing Generated and Captured Data ...
Страница 199: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 199 Using Auxiliary Functions ...
Страница 215: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 215 How Do I ...
Страница 240: ...Index 240 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...