How to Set Up a Port or Channel for Generated Signals
Setting Up Ports and Channels
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
111
–
Hold Width
or
Hold Duty Cycle
: Set a preference if the system clock
frequency is changed.
–
Transition times
(with HP E4842A or E4838A frontends only):
The Rise/Fall Time can be set in the range of 0.5 to 10.0 ns to achieve
the specified 0.7 to 6.0 ns. The rise and fall times are coupled and
yield a signal as shown below:
Figure 38
Variable Slopes Setting
–
Format
: Choose from the list.
Figure 39
Signal Formats
The default format for a data port and the channels connected to a data
port is NRZ. The default format for a pulse port or an unconnected
generator frontend is RZ.
NRZ
RZ
R1
1
0
1
1
Non-return to zero
Return to zero
Return to one
Содержание 81200
Страница 1: ...Revision 2 1 HP 8 1 200 D a t a G e n e r a t o r A n a l y z e r P l a t f o r m User Guide ...
Страница 9: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 9 What s New in This Release ...
Страница 15: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 15 Introduction to the System ...
Страница 49: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 49 Test Development Overview ...
Страница 57: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 57 Structure of the User Interface ...
Страница 83: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 83 Setting Global System Parameters ...
Страница 93: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 93 Connecting the DUT ...
Страница 105: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 105 Setting Up Ports and Channels ...
Страница 127: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 127 Choosing the Kind of Measurement ...
Страница 155: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 155 Creating and Editing Segments ...
Страница 171: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 171 Using the Data Sequence Editor ...
Страница 181: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 181 Running the Test ...
Страница 186: ...Running the Test How to Start Stop the Test 186 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...
Страница 187: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 187 Viewing Generated and Captured Data ...
Страница 199: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 199 Using Auxiliary Functions ...
Страница 215: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 215 How Do I ...
Страница 240: ...Index 240 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...