Hardware
W
arning
Errors
6000
{
6999
Hardware-warning
errors
occur
when
the
hardware
is
faulty
.
The
spectrum
analyzer
can
still
make
measurements
,
but
the
accuracy
of
the
measurement
cannot
be
guaranteed.
6000
EAROM
unprotected
Cause:
The
memory-enable
write
switch
is
set
to
the
WRITE
position.
It
is
not
set
to
the
PROTECT
position.
Cure:
Set
the
memory-enable
write
switch
to
the
PROTECT
position.
6001
Confidence
test
passed
6002
A6
RAM
checksum
(battery)
6003
FFS
won't
tune
low
6004
FFS
won't
tune
high
6005
Idler
tuning
range
6006
YTO
tuning
range
6007
MSIB
NMAA
received
HP-MSIB
\no
module
at
address"
received
Cause:
The
module
attempted
to
establish
communication
with
a
module
at
an
incorrect
address
.
The
system
responded
with
an
NMAA
(no
module
at
address).
Cure:
If
using
MSIB
communication,
verify
the
address
if
the
module
you
are
trying
to
communicate
with
is
correct.
Otherwise
,
document
all
steps
that
led
to
the
error
condition
and
contact
your
HP
representative
.
6-22
Error
Messages
Содержание 70000 series
Страница 42: ...Auxiliary Measurement Indicators Figure 1 8 Auxiliary Measurement Indicators Overview 1 19 ...
Страница 67: ......
Страница 95: ......
Страница 346: ...Figure A 2 Simpli ed Lightwave Signal Analyzer Block Diagram Theory of Operation A 3 ...
Страница 366: ...Figure B 6 HP 70620A B Preampli er and HP 70621A Preampli er Front View Input and Output Ports B 11 ...
Страница 395: ......
Страница 447: ......