Current
Error
Indicator
LED
.
The
CURRENT
error
indicator
lights
when
the
mainframe
power
supply
senses
overloading.
1.
Remove
one
module
from
the
mainframe
.
2.
Cycle
the
power
.
Is
the
CURRENT
indicator
lit?
If
yes:
The
module
is
not
faulty
.
Repeat
Steps
1
and
2
until
a
faulty
module
is
identied.
If
no:
The
module
is
probably
faulty
.
T
o
verify
,
replace
the
module
in
the
mainframe
and
check
if
the
CURRENT
indicator
lights
again.
F
or
repair
alternatives
,
refer
to
\Repair
Procedures".
3.
If
the
CURRENT
indicator
is
still
lit
when
all
of
the
modules
have
been
removed
from
the
mainframe
,
the
mainframe
is
probably
faulty
.
Refer
to
the
HP
70001A
Service
Guide
for
further
troubleshooting
information.
I/O
Check
Error
Indicator
LED
.
The
I/O
CHECK
LED
should
be
o
when
both
HP-MSIB
cables
are
either
connected
to
or
disconnected
from
the
HP
70001A
mainframe
.
The
LED
should
be
on
if
one
of
the
HP-MSIB
cables
is
disconnected
from
the
mainframe
.
It
also
comes
on
when
any
mainframe
or
display
on
HP-MSIB
is
not
powered
on.
If
the
indicator
is
on,
no
element
can
power
up
until
it
is
cleared.
The
following
procedure
may
be
used
to
isolate
reasons
for
the
I/O
CHECK
LED
to
be
on.
1.
V
erify
that
the
power
to
all
mainframes
and
stand-alone
display
instruments
on
the
HP-MSIB
is
on.
2.
Check
that
all
HP-MSIB
cables
are
properly
and
securely
connected.
3.
Disconnect
the
HP-MSIB
cables
from
the
mainframe
.
Is
the
I/O
CHECK
LED
still
lit?
If
yes:
The
HP
70001A
mainframe
is
probably
faulty
.
Refer
to
the
HP
70001A
Service
Guide
for
further
troubleshooting
information.
If
no:
The
problem
is
either
the
cables
or
the
instrument
that
was
connected
to
the
display
with
the
cables
.
1.
First
test
both
cables
at
the
same
time
.
With
one
end
of
each
cable
attached
to
the
\IN"
and
\OUT"
HP-MSIB
mainframe
connectors
,
connect
the
opposite
ends
of
the
two
cables
together
.
2.
If
the
I/O
mainframe
light
comes
on,
one
of
the
two
cables
is
faulty
.
3.
Remove
one
of
the
cables
.
Connect
the
remaining
cable
so
that
it
is
connected
to
the
\IN"
and
\OUT"
HP-MSIB
mainframe
connectors
.
4.
If
the
I/O
mainframe
light
comes
on,
the
cable
is
faulty
.
If
the
I/O
light
does
not
come
on,
test
the
other
cable
.
5.
If
the
light
does
not
come
on
for
either
cable
,
then
the
element
connected
with
these
cables
is
faulty
.
Refer
to
the
service
manual
for
that
element
for
troubleshooting
information.
An
alternate
troubleshooting
approach
is
to
reconnect
the
HP-MSIB
cables
and
use
the
HP-MSIB
Troubleshooting
Utility
to
isolate
the
problem
further
.
Note
Any
elements
responding
to
the
HP-MSIB
Troubleshooting
Utility
with
COMMUNICATION
COMPLETE
indicate
that
the
cables
are
not
faulty
.
5-24
T
roubleshooting
Содержание 70000 series
Страница 42: ...Auxiliary Measurement Indicators Figure 1 8 Auxiliary Measurement Indicators Overview 1 19 ...
Страница 67: ......
Страница 95: ......
Страница 346: ...Figure A 2 Simpli ed Lightwave Signal Analyzer Block Diagram Theory of Operation A 3 ...
Страница 366: ...Figure B 6 HP 70620A B Preampli er and HP 70621A Preampli er Front View Input and Output Ports B 11 ...
Страница 395: ......
Страница 447: ......