![HP 70000 series Скачать руководство пользователя страница 305](http://html.mh-extra.com/html/hp/70000-series/70000-series_user-manual_161488305.webp)
The
\Module
Removal
Method"
may
only
be
used
to
troubleshoot
the
HP-MSIB
when
error
indicators
are
ashing.
Modules
with
ashing
error
indicators
are
individually
removed
from
the
HP
70001A
mainframe
or
HP
70004A
color
display
until
the
faulty
module
or
HP
70001A
mainframe/HP
70004A
color
display
is
isolated.
This
method
involves
removal
and
reinstallation
of
rear
panel
cables
.
If
none
of
the
elements
on
the
HP-MSIB
can
communicate
,
either
a
module
is
disrupting
the
entire
HP-MSIB
,
the
HP
70001A
mainframe
or
HP
70004A
color
display
HP-MSIB
circuits
are
faulty
,
or
the
display
element
has
a
faulty
HP-MSIB
interface
and
the
utility
is
giving
erroneous
results
.
Use
the
following
steps
to
further
isolate
the
problem.
1.
If
an
abnormal
screen
display
is
present,
refer
to
the
HP
70001A
Service
Guide
or
HP
70004A
Service
Guide
and
to
the
appropriate
display
element
service
manual
for
further
troubleshooting
and
repair
information.
2.
If
ashing
error
indicators
occur
,
the
faulty
element
or
HP
70001A
mainframe/HP
70004A
color
display
may
be
isolated
by
using
the
steps
given
below
.
a.
If
a
module
is
under
test,
turn
the
HP
70001A
mainframe
or
HP
70004A
color
display
OFF
.
b.
Remove
the
element
from
the
HP-MSIB
by
removing
the
module
or
by
disconnecting
the
HP-MSIB
cables
from
a
stand-alone
element.
c.
Turn
the
system
power
o,
then
on
again.
d.
If
the
problem
is
gone
,
the
element
is
faulty
.
Refer
to
the
appropriate
service
manual
or
technical
reference
manual
for
further
troubleshooting
and
repair
information.
e.
If
the
problem
remains
,
repeat
Steps
a
through
d
on
each
module
until
the
faulty
element
has
been
isolated.
If
none
of
the
elements
is
faulty
,
the
HP
70001A
mainframe
or
HP
70004A
color
display
is
faulty
.
Refer
to
the
HP
70001A
Service
Guide
or
HP
70004A
Service
Guide
for
further
troubleshooting
and
repair
information.
Module
Removal
Method
This
method
involves
removing
one
module
at
a
time
until
the
faulty
module
,
HP
70001A
mainframe
or
HP
70004A
color
display
is
isolated.
Because
it
uses
ashing
error
lights
to
isolate
the
problem,
this
method
cannot
be
used
to
troubleshoot
when
the
only
symptom
is
an
abnormal
screen
display
.
If
the
error
indicator
of
only
one
element
ashes
at
a
1
Hz
rate
,
that
element
cannot
communicate
over
the
HP-MSIB
and
is
probably
faulty
.
If
the
error
indicators
of
more
than
one
module
are
ashing,
either
the
HP
70001A
mainframe
or
HP
70004A
color
display
HP-MSIB
is
faulty
or
a
faulty
module
is
disrupting
the
entire
HP-MSIB
.
It
is
possible
for
a
module
to
disrupt
the
entire
HP-MSIB
without
having
its
error
indicator
ash.
Remove
the
modules
that
have
ashing
error
indicators
one
at
a
time
.
If
all
of
the
error
indicators
stop
ashing
when
a
module
is
removed,
that
module
is
faulty
.
F
or
additional
troubleshooting
and
repair
information,
refer
to
\Repair
Procedures".
If
no
modules
appear
to
be
faulty
,
the
problem
is
with
the
HP
70001A
mainframe
or
HP
70004A
color
display
.
Refer
to
the
HP
70001A
Service
Guide
or
HP
70004A
Service
Guide
for
further
troubleshooting
and
repair
information.
F
or
information
on
reconnecting
the
modules
after
troubleshooting,
see
the
HP
70000
Modular
Spectrum
Analyzer
Installation
and
V
erication
Manual.
5-26
T
roubleshooting
Содержание 70000 series
Страница 42: ...Auxiliary Measurement Indicators Figure 1 8 Auxiliary Measurement Indicators Overview 1 19 ...
Страница 67: ......
Страница 95: ......
Страница 346: ...Figure A 2 Simpli ed Lightwave Signal Analyzer Block Diagram Theory of Operation A 3 ...
Страница 366: ...Figure B 6 HP 70620A B Preampli er and HP 70621A Preampli er Front View Input and Output Ports B 11 ...
Страница 395: ......
Страница 447: ......