![HP 70000 series Скачать руководство пользователя страница 313](http://html.mh-extra.com/html/hp/70000-series/70000-series_user-manual_161488313.webp)
User
Application
Errors
0001
{
0999
0027
Res
BW-amplitude
Cause:
The
amplitude
variation
between
the
resolution
bandwidths
of
the
HP
70902A
IF
section
is
too
large
.
Cure:
The
bandwidth
lter
adjustments
in
the
HP
70902A
IF
section
should
be
performed.
If
this
does
not
remedy
the
problem,
repair
the
HP
70902A
IF
section
LC
or
crystal
bandwidth
lters
.
0028
Calibration
atten
Cause:
The
calibration
attenuator
of
the
HP
70902A
IF
section
is
not
functioning
correctly
.
Cure:
The
calibration
attenuators
in
the
HP
70902A
IF
section
require
adjustment
or
repair
.
0029
Log
amplifier
Cause:
The
log
amplier
of
the
HP
70902A
IF
section
is
not
functioning
correctly
.
Cure:
P
erform
the
HP
70902A
IF
section
log
amplier
adjustment.
If
this
does
not
remedy
the
problem,
refer
to
the
HP
70902A
IF
section
Service
Manual
for
procedures
to
repair
the
log
amplier
or
replace
the
thick-lm
log
amplier
microcircuits
.
*0030
21.4MHz
input
cable
Cause:
The
program
cannot
nd
the
rear
panel
21.4
MHz
input
signal
for
the
HP
70903A
IF
section.
Cure:
Check
the
21.4
MHz
OUT
connection
on
the
rear
panel
of
the
RF
section
going
to
the
21.4
MHz
IN
connection
on
the
rear
panel
of
the
HP
70903A
IF
section.
If
this
cable
is
connected
properly
,
it
might
be
defective
.
*0031
Video
output
cable
Cause:
The
rear
panel
VIDEO
OUT
signal
of
the
HP
70903A
IF
section
is
not
getting
to
the
HP
70900B
local
oscillator
source
VIDEO
input.
Cure:
Check
the
rear
panel
VIDEO
OUT
connection
on
the
HP
70903A
IF
section.
Be
sure
it
is
routed
properly
to
the
HP
70900B
local
oscillator
source
and
the
cable
is
not
defective
.
0032
LC
board
Cause:
The
HP
70903A
IF
section
LC
board
signal
amplitude
is
too
low
.
Cure:
P
erform
the
HP
70903A
IF
section
LC
lter
adjustments
or
the
HP
70903A
IF
section
step
gain
adjustment.
If
these
do
not
remedy
the
problem,
the
A1
LC
lter
assembly
requires
repair
.
6-6
Error
Messages
Содержание 70000 series
Страница 42: ...Auxiliary Measurement Indicators Figure 1 8 Auxiliary Measurement Indicators Overview 1 19 ...
Страница 67: ......
Страница 95: ......
Страница 346: ...Figure A 2 Simpli ed Lightwave Signal Analyzer Block Diagram Theory of Operation A 3 ...
Страница 366: ...Figure B 6 HP 70620A B Preampli er and HP 70621A Preampli er Front View Input and Output Ports B 11 ...
Страница 395: ......
Страница 447: ......