SVC 5- 23
Replace Cryogenic Valve and/or Nozzle
HAZARDOUS VOLTAGES ARE PRESENT IN THE INSTRUMENT WHEN THE POWER
CORD IS CONNECTED. AVOID A POTENTIALLY DANGEROUS SHOCK HAZARD BY
DISCONNECTING THE POWER CORD BEFORE WORKING ON THE INSTRUMENT.
1. Set the main power line switch to the off position.
2. Disconnect the power cable from its receptacle.
3. Allow time for the oven and heated zones to cool.
4. When the heated zones are cool, turn off all gas
supplies.
5. Remove the electronics carrier top cover by
grasping it at the rear and lifting upwards until
its catch releases, then pulling it towards the
rear of the instrument.
6. Remove the right side panel by removing four screws: two each along its upper and lower
edges.
7. Remove the two screws securing the left side panel along its bottom edge.
8. Slide the left side panel towards the rear of the instrument and lift.
9. Remove the four screws securing the rear cover to the instrument.
OFF
ON
RED O VISIBLE
SCREWS
ELECTRONICS
CARRIER TOP
COVER
SCREWS
SCREWS
RIGHT SIDE
PANEL
SCREWS
SCREWS