SVC 6- 8
22. Use the lance release tip of an AMP pin extrac-
tion/lance reset tool (8710- 1542) to remove the
appropriate pins from connector P7. (The tool
features a lance release tip and a lance reset
tip. The lance release tip is used to depress
the pin locking lance to extract the pin from a
connector. The lance reset tip positions a lock-
ing lance to its proper height to ensure reten-
tion of the pin in the connector.)
23. Prepare the pins corresponding to the replace-
ment sensor cartridges by adjusting their lock-
ing lances using the lance reset portion of the
tool.
24. Insert the pins for the replacement sensor into
their appropriate locations in the plug, making
sure the locking lance on each pin seats into its
hole through the side of the plug.
25. Gently pull on the wire to ensure that the pin is locked in the connector.
26. Insert connector P7 into its corresponding receptacle on the main PCB.
27. After the leads have been exchanged in the appropriate connectors, proceed to the heater/sen-
sor cable assembly removal/replacement procedure applicable to the desired heated zone.
EXTRACTION
LANCE RESET
TOOL
LANCE
RELEASE
TIP
TOOL
STOP
DASH NUMBER
MARKED HERE
LANCE
RESET TIP
TOOL STOP
INSERT TIP
HERE
(BACK OF
CAVITY)
LANCE
RELEASE
TIP
LOCKING
LANCE
CAVITY
BACK OF
HOUSING
CONTACT
(TYPICAL)
RECEPTACLE
BOX (TYPICAL)
LOCKING
LANCE
TAPER
LANCE
RESET
TIP
LOCKING
LANCE
CAVITY
LOCKING
LANCE
BACK OF
HOUSING
CONTACT
(TYPICAL)