SVC 4- 29
`
When disconnecting a plug, pull on the
plug not on its wires. Pulling on the wires
may cause breakage.
`
The following steps require protection
against ESD (Electro- Static Discharge).
Use a grounded wrist strap (part no.
9300- 0969 - large, or 9300- 0970 -
small) connected to a suitable ground.
Failure to heed this caution may result in
damage to the instrument.
9. Disconnect the NPD bead power cable from the
detector PCB by pulling it straight off.
10. Use the side edge of an AMP pin extraction/
lance tool (8710- 1542) to remove the interlock-
ing side covers from the plug. (The plug is
made up of two connectors; one is for the to-
roid bead power, the other is for the active con-
trol element.)
11. Loosen the screws securing the detector PCB
interconnect clamps to the thermal strap.
DETECTOR COVER
COLLECTOR
INTERCONNECT
CLAMP SCREW
TO NPD TRANSFORMER
TOP PART OF CONNECTOR
TO NPD BEAD
POWER
CONTROL
BOTTOM PART
OF
CONNECTOR
LOCKING TAB
LOCKING TAB
NOTE
ASSEMBLED
CONNECTORS
5
1
5
1
EXTRACTION
LANCE RESET
TOOL
LANCE
RELEASE
TIP
TOOL
STOP
DASH NUMBER
MARKED HERE
LANCE
RESET TIP