SVC 2- 17
9. Disconnect the Carrier Gas Inlet which termi-
nates either at the Forward Pressure Regulator
(for Manual Pressure Control) or the PCOC Pro-
portional Control Valve (for Electronic Pressure
Control).
10. Disconnect the Septum Purge Outlet which ter-
minates at the PCOC Purge Regulator,
mounted inside the flow module.
11. Cap the base of the inlet, inside the column
oven.
12. Remove the auto- injection assembly (or op-
tional manual injection assembly) by rotating it
counter- clockwise). Be careful not to loose
the septum, insert, or PCOC insert spring which
are installed under the injection assembly.
AUTO
INJECTION
ASSEMBLY
OPTIONAL
MANUAL
INJECTION
ASSEMBLY
DUCKBILL
SEPTUM
SEPTUM
PCOC INSERT
SPRING
AIR DEFLECTOR
WELDMENT
INSULATION
CAVITY
SLEEVE
SCREW
INSERT
HEATER/SENSOR
CABLE ASSEMBLY
13. Remove the air deflector using a Pozidriv screwdriver to remove the screw securing the air de-
flector to the inlet weldment.