Extended
With the extended interface, both the software and the hardware can
Interface with
Hardware
Handshake
control the data flow between the HP
and the controller. This
allows you to have more control of data flow between devices. The
HP
uses the following connections on its RS-232C interface for
extended interface communication:
l
Pm 7 SGND (Signal Ground)
l
Pin 2 TD (Transmit Data from HP
l
Pin 3 RD (Receive Data into HP
The additional lines you use depends on your controller’s implementation
of the extended
interface.
l
Pin 4 RTS (Request To Send) is an output from the
HP
which can be used to control incoming data flow.
l
Pm 5 CTS (Clear To Send) is an input to the HP
which controls data flow from the HP
l
Pm 6 DSR (Data Set Ready) is an input to the HP
which controls data flow from the HP
within two bytes.
l
Pin 8 DCD (Data Carrier Detect) is an input to the
HP
which controls data flow from the HP
within two bytes.
l
Pin 20 DTR (Data Terminal Ready) is an output from the
HP
which is enabled as long as the HP
is
turned on.
The TD (Transmit Data) line from the HP
must connect to the
RD (Receive Data) line on the controller. Likewise, the RD
from the
HP
must connect to the TD line on the controller.
HP
16518
Programming Reference
Programming Over RS-232C
3-3
Содержание 1650B
Страница 2: ......
Страница 16: ......
Страница 40: ......
Страница 48: ......
Страница 58: ......
Страница 74: ......
Страница 76: ...System Commands 6 2 Figure 6 l System Commands Syntax Diagram HP 1650B HP 1651B Programming Reference ...
Страница 112: ......
Страница 128: ......
Страница 134: ......
Страница 160: ......
Страница 184: ... L h E I Figure 13 1 SLlSt Subsystem Syntax Diagram SLIST Subsystem 13 2 HP 165081HP Is5 B Programming Reference ...
Страница 185: ...Figure 13 l SLlSt Subsystem Syntax Diagram continued HP 1650B HP 16518 Programming Reference SLIST Subsystem 13 3 ...
Страница 210: ......
Страница 236: ......
Страница 282: ......
Страница 318: ......
Страница 336: ......
Страница 346: ......