RUNTil
RUNTil
command/query
The RUNTil (run until) command allows you to
a stop condition
when the trace mode is repetitive. Specifying
causes the analyzer to
run until either the front-panel STOP key is pressed or the STOP
command is issued.
There are four conditions based on the time between the X and 0
markers. Using this difference in the condition is effective only when time
tags have been turned on (see the TAG command in the
subsystem). These four conditions are as follows:
l
The difference is less than (LT) some value.
l
The difference is greater than (GT) some value.
l
The difference is inside some range
l
The difference is outside some range
End points for the
and
should be at least 10
apart.
There are two conditions which are based on a comparison of the
acquired state data and the compare data image. You can run until one of
the following conditions is true:
l
Compare equal
Every channel of every label has the
same value.
l
Compare not equal
Any channel of any label has a
different value.
The RUNTil query returns the current stop criteria for the comparison
when running in repetitive trace mode.
The RUNT’1 instruction (for state analysis) is available in both the
and
subsystems.
Subsystem
HP
1651B
Programming Reference
Содержание 1650B
Страница 2: ......
Страница 16: ......
Страница 40: ......
Страница 48: ......
Страница 58: ......
Страница 74: ......
Страница 76: ...System Commands 6 2 Figure 6 l System Commands Syntax Diagram HP 1650B HP 1651B Programming Reference ...
Страница 112: ......
Страница 128: ......
Страница 134: ......
Страница 160: ......
Страница 184: ... L h E I Figure 13 1 SLlSt Subsystem Syntax Diagram SLIST Subsystem 13 2 HP 165081HP Is5 B Programming Reference ...
Страница 185: ...Figure 13 l SLlSt Subsystem Syntax Diagram continued HP 1650B HP 16518 Programming Reference SLIST Subsystem 13 3 ...
Страница 210: ......
Страница 236: ......
Страница 282: ......
Страница 318: ......
Страница 336: ......
Страница 346: ......