User Manual: HotDog

TM

 Veterinary Controller                                                                                  Page 12   

 

Error 

Codes 

Error Description 

Problem Solving Steps 

E1 

Over-Temperature 

When the temperature exceeds one degree above 
the set point, the Alarm sounds and power is 
removed from the Warming Blanket.  Unplug the 
Blanket to reset the Alarm.  Wait 5 minutes and 
then reconnect the Blanket.  Turn the Controller 
on and select the temperature.  If the Alarm 
occurs again, stop using the Blanket and contact 
technical support.   

E2 

Failure To Reach Temp 

When the Warming Blanket does not reach the 
temperature set point within 10 minutes, the 
Alarm sounds, and power is removed from the 
Blanket.  Check to make sure the Blanket is in 
contact with a patient and the sensor area is 
touching the patient.  Unplug the Blanket and 
reconnect to reset.  If the Alarm occurs again, 
stop using the Blanket and contact technical 
support.   

E3 

Port Current Reached 

If the electrical current in the Warming Blanket 
exceeds the allowable limit, the Alarm  sounds 
and power is removed from the Blanket.  This 
may indicate and electrical problem with the 
Blanket.  Unplug the Blanket from the Controller 
and reconnect to reset.  If the Alarm occurs again, 
stop using the Blanket and contact technical 
support.   

E4 

Sensor or Cable Failure 

If the Controller loses communication with the 
sensor in the Warming Blanket, an Alarm sounds, 
and power is removed from the Blanket.  This 
may be caused by an electrical problem in the 
Blanket or in the Controller.  Swap the cables and 
Blanket with known good product to isolate the 
problem if possible.  If the problem continues, 
stop using the Blanket or cable and contact 
technical support  

E8 

Over-Temperature Sensor 

(Secondary) 

The temperature sensor has exceeded 46° C.  
Disconnect the Warming Blanket and contact 
technical support. 

EA, EC, 

EF, EH or 

EP 

Hardware Failure 

Please turn off the Controller, wait one minute, 
then restart.  If the problem persists contact 
technical support. 

 

Contact Technical Support @ 952.465.3500 or 

[email protected] 

 
 
 
 
 
 

 

Содержание WC52-R(VET)

Страница 1: ...Warming System Controller Model WC52 R VET User Manual www vetwarming com Manufactured by Augustine Temperature Management LLC 6581 City West Parkway Eden Prairie MN 55344 USA TEL 952 465 3500 FAX 952 465 3501 www vetwarming com ...

Страница 2: ...Initial Setup Assembly 4 Contents 4 Mounting the HotDog Controller to an IV Pole 5 Control Panel Features Operating Modes 6 Instructions for Use 8 Pressure Reduction 9 Dental Procedures 9 Cleaning the Controller 9 Frequently Asked Questions 10 Error Codes Alerts and Alarms 11 Definition of Product Symbols 13 Specifications 13 Electromagnetic Compatibility EMC 15 ...

Страница 3: ... that warming is not advisable INDICATIONS FOR USE The System is intended to prevent or treat hypothermia and provide warmth to veterinary patients The System should be used in circumstances in which patients may not be able to maintain a state of normothermia CONTRAINDICATIONS Do not warm ischemic or non perfused tissue thermal injury may result Examples include tissue distal to aortic cross clam...

Страница 4: ...does not pose any potential hazard Veterinary product only for use with animals Do not continue to use the System if a safety Alarm sounds after reset See Page 11 12 Error Codes Alerts Alarms for detailed troubleshooting information Do not open the Controller There are no user serviceable parts If service is required contact Customer Service The limited warranty is invalidated if The Controller is...

Страница 5: ...dle counterclockwise until the unit releases Caution To prevent the IV pole from tipping the Controller must be attached at a height that provides stability It is recommended to use an IV pole with a minimum wheelbase radius of 35 6 cm 14 in and to mount the Controller no higher than 112 cm 44 in from the floor Failure to properly mount the Controller may result in IV pole tipping catheter site tr...

Страница 6: ...ratures will be displayed in Celsius 4 Remove the Blanket from the shipping carton The serial number and 2 year expiration date are written on the yellow cable extending from the Blanket 5 Connect the yellow cable to the short yellow cable off the side of the Blanket by aligning both red dots and pushing straight in Do not twist 6 Position and secure the Blanket 7 Insert the Blanket s connecting c...

Страница 7: ...nnected and the port icon illuminates on the screen Note The C D and E ports are not available for Veterinary use Temperature Setting Display When a warming blanket is plugged into a port an audible beep will sound and the display will show two dashes Press the Temperature Adjust Button for the desired port until the desired temperature is displayed The temperature can be selected in one degree in...

Страница 8: ... a yellow port with the red dot on the cable facing up 6 Turn the Controller on to begin warming and select the desired temperature on the Controller The Blanket will take approximately 3 minutes to reach the selected temperature 7 Monitor the temperature of the patient frequently during each case Increase or decrease the Blanket temperature gradually to maintain normothermia GUIDELINES FOR TEMPER...

Страница 9: ...ps are taken to mitigate pressure Do not place the cautery plate on the sensor area of the Blanket DENTAL PROCEDURES Although the System is particularly useful in dental procedures clinicians should take a few extra precautions to protect the patient Most importantly ensure that the patient is in contact with the sensor at all times during the case Follow the temperature selection and pressure red...

Страница 10: ...istance and the time it needs to reach temperature When new it only takes a few minutes to reach the set temperature After approximately24 months of use the Blanket will take longer to reach the set temperature We have no data to support the use of the devices beyond 24 months 4 Are the Blankets difficult to clean Cleaning takes 30 seconds or less Blankets can be cleaned by wiping with an approved...

Страница 11: ...act with the sensor on the black side of the Banket Use the Blanket strap to secure 12 Can the HotDog system be used in x ray Yes The heater fabric is completely radiolucent However the Blanket has parallel busbars that run along the long edges These can be seen on x ray The area around the sensor is also radio opaque Position the Blanket so that the imaging area does not contain the sensor or bus...

Страница 12: ...e electrical current in the Warming Blanket exceeds the allowable limit the Alarm sounds and power is removed from the Blanket This may indicate and electrical problem with the Blanket Unplug the Blanket from the Controller and reconnect to reset If the Alarm occurs again stop using the Blanket and contact technical support E4 Sensor or Cable Failure If the Controller loses communication with the ...

Страница 13: ...nge Fuse Equipotential CE Mark IPX2 Protected against dripping water when tilted up to 15 vertically dripping water shall have no harmful effect when the enclosure is tilted at an angle up to 15 from its normal position Equipment Classified by Intertek Testing Services NA Inc with respect to electric shock fire and mechanical hazards only in accordance with UL 60601 1 SPECIFICATIONS Physical Chara...

Страница 14: ...roller will discontinue power to warming blanket Over current limits Port A 10 amps max Port B 10 amps max Port C 5 amps System 14 6 amps System Over current Protection Dual input fused lines Electrical Characteristics Leakage Current Meets UL 2601 1 and IEC 60601 1 requirements for Class I Type BF equipment Power Consumption 850W maximum Power Cord 4 6 m 15 ft Device Ratings Input 100 240 VAC 50 ...

Страница 15: ...ecautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this user manual Warning Accessories and cables other than those specified may result in increased emissions or decreased immunity of the HotDog Veterinary Patient Warming system The HotDog Veterinary Patient Warming System should not be used adjacent to or stacked with other equipme...

Страница 16: ...line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip ...

Страница 17: ...RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and refl...

Страница 18: ... applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures obje...

Отзывы: