User Manual: HotDog
TM
Veterinary Controller Page 18
Recommended separation distances between portable and mobile RF communications
equipment and the HotDog Veterinary Patient Warming System
The HotDog Veterinary Patient Warming System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the HotDog Veterinary Patient Warming System can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the HotDog Veterinary Patient Warming System as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
P
35
,
0
d
800 MHz to 2,5 GHz
P
7
,
0
d
0,01
0,12
0,04
0,07
0,1
0,37
0,11
0,22
1
1,2
0,35
0,70
10
3,7
1,1
2,2
100
12
3,5
7,0
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
TECHNICAL SUPPORT & CUSTOMER SERVICE
Please have the serial number of your HotDog Controller when you call for technical support. The serial
number is located on the back of the Controller.
USA and Worldwide:
Augustine Temperature Management, LLC
6581 City West Parkway
Eden Prairie, MN 55344 USA
TEL 952.465.3500
FAX 952.465.3501
www.vetwarming.com
Customer Service Contact
Phone: +1-952-465-3500 Email:
HotDog is a trademark of Augustine Temperature Management, registered in the U.S. Patent & Trademark Office. Devices are protected by
some or all of the following patents: (US Patents 7,543,344; 7,714,255; 7,851,729; 7,786,408; 8,062,343; 8,283,602; 8,604,391; 8,624,164;
8,772,676; 8,986,359; 9,962,122; 9,668,303; 10,154,543; 10,201,935; 10,206,248; 10,506,668; PCT Patent EP 2,062,460) . Other patents are
pending.
© 2022 Augustine Surgical, Inc. All rights reserved.
2836 Rev D (10/2022)