User Manual: HotDog
TM
Veterinary Controller Page 3
INTRODUCTION
Device Description
The HotDog Veterinary Patient Warming System consists of a Controller, and a reusable Warming
Blanket. It is a veterinary product for use only with animals.
Controller:
Allows you to select the temperature for your Warming Blankets. The WC52R (VET)
operates 2 Blankets.
Warming Blankets
:
Blankets are available in several sizes; select the proper size.
Model
Number
Warming Blanket Size
Primary Patient Size
V101
X-Small 10” X 12” (25.4 x 30.5 cm)
Mice and rats in research
V102
Small 10” X 23” (25.4 x 58.4 cm)
Exotics and Pocket pets
V103
Medium 16” X 22” (40.6 x 55.9 cm)
An average-size cat
V104
Large 22” X 31” (55.9 x 78.7 cm)
An average-size dog
V106
Extra Large 27” X 47” (68.6 x 119.4 cm)
Bigger Dogs
It is the responsibility of the clinician to determine whether warming is appropriate for each patient.
The System should not be used when clinical considerations indicate that warming is not advisable.
INDICATIONS FOR USE
The System is intended to prevent or treat hypothermia and provide warmth to veterinary patients.
The System should be used in circumstances in which patients may not be able to maintain a state
of normothermia.
CONTRAINDICATIONS
Do not warm ischemic or non-perfused tissue; thermal injury may result. Examples include
tissue distal to aortic cross clamping, severe dehydration, or when vasoconstrictive drugs
would lead to severe, prolonged vasoconstriction.
Do not warm patients receiving transdermal medication as increased drug delivery may
occur.
WARNINGS
Explosion hazard
- Do not use in the presence of flammable anesthetics or highly oxygen-
enriched environments such as hyperbaric chambers, oxygen tents, etc.
Inspect the Warming Blanket prior to use and cleaning for signs of damage or excessive wear
such as cuts, holes, or loose electrical connections. If signs of wear, nicks, or cuts are
evident, remove the Blanket from service.
Discontinue use of the System if the over-temperature indicator and/or any other Alarm
continues to sound after reset. See Error Codes: Alerts and Alarms section.