User Manual: HotDog
TM
Veterinary Controller Page 13
DEFINITION OF PRODUCT SYMBOLS
See IFU for Warnings and
Precautions
See information for
use at the provided
website.
Temperature Sensor
Natural Rubber Latex
Free
Not Sterile
Alarm
Mains Power On
Indicator
Temperature
Adjustment
Indicates separate treatment
from general waste at end of
life. See Precautions for details.
Attention; consult
accompanying
documents
Reference Number
Lot Number
BF Patient Applied Part
according to IEC60601-1.
Serial Number
Manufacture Date
Do not use after
YYYY-MM-DD
Transport and storage
temperature range
Manufacturer
Keep Dry
Transport and storage
humidity range
Fuse
Equipotential
CE-Mark
IPX2
Protected against dripping water when tilted up to 15°; vertically dripping water shall have no harmful effect when the
enclosure is tilted at an angle up to 15° from its normal position.
Equipment Classified by Intertek Testing Services NA Inc. with respect to electric shock, fire, and mechanical hazards only, in
accordance with UL 60601-1.
SPECIFICATIONS
Physical Characteristics
Dimensions
33 cm high x 14.0 cm deep x 19.7 cm wide13” high x 5.5” deep x 7.75” wide
Weight
5 kg (11 lbs.)
Mounting
Can be placed on a horizontal flat surface (i.e. table top), clamped to an IV pole or
hung on a OR/gurney rail using optional hanging hooks
Temperature Characteristics
Temperature Control
Micro-processor
Operating Temperatures
Blanket Ports A and B adjustable in 1
o
C increments
37° to 43° ± 1.0°C
98.6° to 109.4° ± 1.8°F
Port C - NA