Chapter 4 Maintenance and Troubleshooting
4.2 Troubleshooting
71
●
ERROR No.0009 Printer communication error
Definition
There is a problem with the printer communication.
Power OFF the instrument and perform the
countermeasures below, and then power ON again.
●
ERROR No.0010 Data memory over
Definition
The number of the memory data exceeds the limit of the
memory.
●
ERROR No.000E Key operation error
Definition
Key operation is invalid.
Cause
Countermeasure
There is a problem
with the printer unit
connection.
Check the printer connection, and con-
nect the instrument and printer again.
The printer is
defective.
Replace the printer.
If the same error occurs again after the
countermeasure is performed, contact our
representatives in your region.
Cause
Countermeasure
Memory over
Check the contents of the memory data,
and delete unnecessary data.
Cause
Countermeasure
Loose connection or
breaking of keys
Disconnect and reconnect the AC
adapter, and press the ON/OFF key 10
seconds later.
If the same error occurs again after the
countermeasure is performed, contact our
representatives in your region.