Chapter 3 Basic Operation
3.2 Settings
19
ATC
In Automatic Temperature Compensation (ATC), the
instrument detects the solution temperature. When the
temperature sensor is connected, the current tempera-
ture of the solutions is displayed automatically. If the
sensor is not connected, the set value for MTC (default:
25
°
C) is displayed.
MTC
In Manual Temperature Compensation (MTC), the tem-
perature sensor of the electrode is not used, and the
solution temperature is entered manually.