Chapter 3 Basic Operation
3.4 Measurement
52
3.
To stop the auto data memory, press the ENTER
key.
・
Do not power ON/OFF while the auto data memory function
works.
It may damage the reliability of the memory data.
・
Only the POWER, ENTER, and the
▲
or
▼
keys are possible
while the auto data memory function works.
・
The actual date of memory may have the difference of ±1
second from the time calculated based on the interval setting.
・
When the number of memory data exceeds 999, data memory
is stopped and ERROR No.0010 is displayed.
3.4.4 Measurement value printing
You can print out measurement values or the HOLD
value.
When the automatic printout setting is ON, automatic
printout is executed at the following occasions.
・
You perform data memory during measurement.
・
Calibration or check is completed in the calibration
mode.
1.
Check that the printer settings and connection are
correctly. (Refer to "3.1.4 Connecting and setting
the printer" P.14.)
2.
Press the
key in the measurement mode or
HOLD state. The currently displayed measurement
value is printed out. Printing examples are shown
below.
▼