Chapter 3 Basic Operation
3.2 Settings
21
3.2.4 pH calibration interval setting
This setting allows you to ensure calibration at a regular
interval. When the set interval (days) elapses after the
previous calibration, ERROR No.0008 (Calibration cycle
error) is displayed.
1.
Press the SET key on the measurement screen,
and press the
▲
or
▼
key to enter the pH calibra-
tion interval setting screen.
2.
Press the ENTER key to enter the calibration inter-
val change screen.
3.
Press the
or
key to change ON or OFF, and
press the ENTER key.
▲
▼