Chapter 3 Basic Operation
3.5 Data operation
58
When the electrode sensitivity is 90% to 93%, "Check" is dis-
played instead of "Good". When the sensitivity is 85% to 90%,
"bAd" is displayed.
Refer to " Electrode status" P.43.
3.5.5 Calibration history printing
The calibration history can be printed out for each
measurement item.
1.
Check the printer settings and that the printer is
connected correctly. (Refer to "3.1.4 Connecting
and setting the printer" P.14.)
2.
Refer to "3.5.4 Calibration history display" P.57 to
display the calibration history.
3.
Press the ENTER key.
The displayed measurement data is printed out.
●
pH check data (without check data)
①
⑤
①
Instrument model
②
Serial number
③
Electrode status
④
Asymmetry potential
⑤
Sensitivity
⑥
Calibration data
④
②
③
⑥