Appendix C: Performing Routine Procedures with SynerJY
71
Triggering
Synapse detection systems offer both input and output TTL trigger functions. Triggering
functions are software enabled. Two hardware triggers are available as SMB receptacles
on the back of the controller: one for TTL input (male) and one for TTL output (female).
Triggering can be activated at the start of each experiment or at the start of each
acquisition during the course of one experiment.
To enable triggering using SynerJY:
1.
Start SynerJY and open the
Experiment Setup
screen.
2. Select the
Triggers
tab. The Triggers window will open.
3. From the
Input Trigger
heading, click
Enable
to activate the
Input Trigger
.
4. Select the appropriate
Input Trigger
parameters.
Event
allows the user to specify
whether the Trigger will be enabled once, at the start of the experiment or at the
Содержание JOBIN YVON SYNAPSE
Страница 1: ...SYNAPSE CCD Detection System User Manual Part Number 81100 Revision 2...
Страница 10: ...Preface x...
Страница 14: ...System Description and Specifications 4...
Страница 28: ...Detector System Installation 18...
Страница 38: ...Initial Power up and Operation 28...
Страница 48: ...Temperature Control 38...
Страница 71: ...61 Appendix A Dimensional Drawings Figure 1 Figure 16 Synapse Detector Head Note Dimensions are in inches mm...
Страница 72: ...Appendix A Dimensional Drawings 62 Figure 17 Distance from Focal Plane to CCD Chip...
Страница 73: ...Appendix A Dimensional Drawings 63 Figure 18 Synapse Power Supply Unit...
Страница 74: ...Appendix A Dimensional Drawings 64...
Страница 86: ...Appendix C Performing Routine Procedures with SynerJY 76...
Страница 100: ...Appendix E Accessories 90...
Страница 104: ...Warranty 94...
Страница 113: ...103 Notes...
Страница 114: ...104 Notes...