67
Appendix
C:
Performing Routine Procedures
with SynerJY
CCD Focus and Alignment on the Spectrograph
1.
Attach a spectral line source, such as a mercury lamp, to the instrument entrance
slit. Consult the documentation provided with your lamp for proper mounting
instructions.
Your light source may emit high-intensity ultraviolet,
visible, or infrared light. Exposure to these types of
radiation, even reflected or diffused can result in
serious, and sometimes irreversible, eye and skin
injuries. When using a lamp, do not aim the light
guide at anyone or look directly into to the light guide
or optical ports of the instrument. Always wear
protective goggles and gloves in conjunction with the
light source.
2.
Start SynerJY. In
Experiment Setup
; select
Monos
from the
General
tab.
3.
Enter an entrance slit width of 13
μ
m (.0130 mm) and manually set the height
limiter to 1 mm.
CAUTION
Содержание JOBIN YVON SYNAPSE
Страница 1: ...SYNAPSE CCD Detection System User Manual Part Number 81100 Revision 2...
Страница 10: ...Preface x...
Страница 14: ...System Description and Specifications 4...
Страница 28: ...Detector System Installation 18...
Страница 38: ...Initial Power up and Operation 28...
Страница 48: ...Temperature Control 38...
Страница 71: ...61 Appendix A Dimensional Drawings Figure 1 Figure 16 Synapse Detector Head Note Dimensions are in inches mm...
Страница 72: ...Appendix A Dimensional Drawings 62 Figure 17 Distance from Focal Plane to CCD Chip...
Страница 73: ...Appendix A Dimensional Drawings 63 Figure 18 Synapse Power Supply Unit...
Страница 74: ...Appendix A Dimensional Drawings 64...
Страница 86: ...Appendix C Performing Routine Procedures with SynerJY 76...
Страница 100: ...Appendix E Accessories 90...
Страница 104: ...Warranty 94...
Страница 113: ...103 Notes...
Страница 114: ...104 Notes...