Clean the Optics
Clean Chemcassette optics annually or whenever optics verification error occurs.
Compressed air is required or per the locations PM schedule.
1. Make sure the Analyzer is out of Monitor Mode.
2. Open the Optics Block Gate.
3. Remove the Chemcassette by releasing and pulling out the analyzer.
4. Open the Vertex Edge side panel.
5. Remove tubing (shown in photo) one at the time and blow out with compressed air.
6. Re-secure side panel and reload the Chemcassette.
NOTE
Remove and clean one port at a time to insure proper orientation of tubing. Do not remove
the capillary tubing (microtubes). After cleaning the Optics, the cleaning counter should be
reset to avoid unnecessary maintenance warning due to Optics cleaning due. (See "Clean
the Optics" above for more information.)
CAUTION
Failure to replace and retighten hardware after service can adversely affect instrument
performance and electromagnetic radiation compliance (EMC). Make certain all fasteners are
reinstalled and firmly fastened.
Vertex Edge System
121
User Manual
Содержание Vertex Edge
Страница 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Страница 2: ......
Страница 6: ......
Страница 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Страница 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Страница 46: ......
Страница 59: ...Vertex Edge System 58 User Manual ...
Страница 63: ...Vertex Edge System 62 User Manual ...
Страница 69: ...Vertex Edge System 68 User Manual ...
Страница 91: ...Vertex Edge System 90 User Manual ...
Страница 96: ...Vertex Edge System 95 User Manual ...
Страница 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Страница 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...