Event
ID
Description
Possible Cause
Resolution
125
Possible debris in optics block
Debris in optics block
Clean optics block
126
Abnormal Optics reading
detected
Compensated optic reading automatically. So no further action required. But too
frequently happens, debris in the optic block.
Clean optics block
127
Accelerated Chemcassette Usage Low level background gas below lower detectable limit
Locate source of background gas
128
Ethernet initialization failed
Failed load the driver, Electronic problem
Reboot the analyzers.
Contact Honeywell Analytics Service.
129
File system of Analyzer is
corrupted
File system corrupted
Contact Honeywell Analytics Service
130
Optics blocks have different
software versions
Program update done incorrectly
Reload program to both optics blocks.
131
Unable To Log event data
File system corrupted
Contact Honeywell Analytics Service
132
Software version mismatched
among Analyzers
New analyzer was installed into the Vertex Edge rack that contains a different software
revision than the other analyzers
Upgrade analyzer Software
133
This Chemcassette is nearing its
expiration date.
Chemcassette not used too long
Replace Chemcassette
134
Chemcassette Type Not Matched Chemcassette with wrong gas family installed
Reinstall Configuration Profile
135
Analyzer configuration failed
Analyzer configuration failed
Reinstall Configuration Profile
136
Point configuration failed
Point configuration failed
Check alarm settings and reinstall Configuration Profile
137
Could not start monitoring
Instrument faults not cleared.
Resolve the reported instrument faults and reset all faults
Invalid RFID tag detected
Reinstall Configuration Profile or Replace Chemcassette
All points disabled
Reinstall Configuration Profile or enable runtime disabled points
Invalid Analyzer/Point Configuration
Reinstall Configuration Profile
LIT in progress
Wait until LIT test is complete and restart monitoring mode
Tape Advance Failure
See recoveries for Fault 233
138
Analyzer RTC not set correctly
Low voltage of the coin battery
Replace the coin battery
RTC failure on the analyzer board
Replace Analyzer.
Vertex Edge System
132
User Manual
Содержание Vertex Edge
Страница 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Страница 2: ......
Страница 6: ......
Страница 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Страница 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Страница 46: ......
Страница 59: ...Vertex Edge System 58 User Manual ...
Страница 63: ...Vertex Edge System 62 User Manual ...
Страница 69: ...Vertex Edge System 68 User Manual ...
Страница 91: ...Vertex Edge System 90 User Manual ...
Страница 96: ...Vertex Edge System 95 User Manual ...
Страница 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Страница 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...