9.
Check the vacuum levels for all connected analyzers. If needed, adjust the vacuum level to
a range of 11 to 13inHG (minimum required is 10inHG). The vacuum level is changed by
using the Supply Vacuum Adjust valve as shown below. All analyzers will adjust as the
supply vacuum adjust valve is turned.
10.
Once complete, leave the flow on for all analyzers and move onto the Adjust Flow Rate
section.
NOTE
At higher altitudes with many analyzers installed, and/or a system on a 50Hz mains,
the system vacuum may not be able to achieve 11inHg. In this case, the maximum
achievable sample point flow rate shown may be reduced by up to 20%..