K-factor
K-factor is a feature that allows gases’ cross-sensitivity to be employed to measure a different
gas using gas calibrations of a selected gas. The detector’s readings are modified by the known
relationship between the two gases. For example, a 5ppm concentration of Gas A is seen by the
detector as the same as a 5.8 ppm concentration of Gas B. The K-factor is 1.2 (5.8 ÷ 5 = 1.16 = 1.2
when rounded to one decimal place).
When k-factor is enabled, gas levels will be multiplied by k-factor automatically. Alarm levels
should be checked again after enabling the k-factor.
NOTE
When mg/m3 is selected as a measurement unit, k-factor is not allowed to be ON.
Non-Zero Warning
When this option is ON and non-zero gas concentration is detected, an informative event will be
recorded and non-zero warning status will be reported to DAq. The point with non-zero warning
will blink in green.
Vertex Edge System
57
User Manual
Содержание Vertex Edge
Страница 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Страница 2: ......
Страница 6: ......
Страница 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Страница 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Страница 46: ......
Страница 59: ...Vertex Edge System 58 User Manual ...
Страница 63: ...Vertex Edge System 62 User Manual ...
Страница 69: ...Vertex Edge System 68 User Manual ...
Страница 91: ...Vertex Edge System 90 User Manual ...
Страница 96: ...Vertex Edge System 95 User Manual ...
Страница 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Страница 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...