Maintenance Faults
A maintenance fault indicates the Vertex Edge System requires attention but is continuing to monitor.
Event
ID
Description
Possible Cause
Resolution
101
Flow is 70 cc/min less than
nominal
Excessive point vacuum due to clog or kinked sample line
Check sample line and line filter.
Clean the sample line and replace filter.
Correct sample line issue.
Sample line too long or ID too small
Ensure sample line requirements are with specifications
Poor gate seal
Check nut on the optic block
Contact Honeywell Analytics Service
Supply vacuum insufficient (less than 7inHg)
Plug pneumatic connector in unused slots Exhaust tubing
restricted Service or switch pumps
Condensation
Check internal lines for moisture
Chemcassette thickness variation
Advance the Chemcassette.
Adjust the needle valve to achieve 200cc/min.
Contact Honeywell Analytics Service.
102
Remaining Chemcassette is low
Less than preset reminder on the Chemcassette
Replace the Chemcassette
103
Analyzer out of monitor too long
Analyzer out of monitor
Reset all alarms and faults, and then return analyzer to Monitor
Mode
Out of Monitor time limit too short
Change time limit in Configuration Profile
104
Run time point disable timeout
Alarms were manually bypassed
Restore point to alarm active mode
Run Time Point Disable time limit too short
Change time limit in Configuration Profile
105
Sample pressures greater than
ambient
Point pressure above atmospheric pressure while in idle
Review sample line location.
Confirm atmospheric conditions.
Pressure sensor Calibration error, Defective sensor
Contact Honeywell Analytics Service
106
Flow is 70 cc/min more than
nominal
Chemcassette thickness variation
Advance the Chemcassette.
Adjust the needle valve to achieve 200cc/min.
Contact Honeywell Analytics Service.
Defective needle valve
Replace needle valve
107
Chemcassette expired
Chemcassette expiration date reached
Replace Chemcassette
Chemcassette installed past its expiration date
Replace Chemcassette
108
No polls from HMI
Communications to HMI PC interrupted longer than 10 seconds
Check Ethernet cable at rear of analyzer.
Check Ethernet hub connection and operation.
Check Ethernet connection to HMI PC.
Vertex Edge System
130
User Manual
Содержание Vertex Edge
Страница 1: ...Vertex Edge 72 POINT CONTINUOUS MONITOR ...
Страница 2: ......
Страница 6: ......
Страница 19: ...Chemcassette 1 Chemcassette directional flow Vertex Edge System 18 User Manual ...
Страница 29: ...Installation Drawings Vertex Edge System 28 User Manual ...
Страница 46: ......
Страница 59: ...Vertex Edge System 58 User Manual ...
Страница 63: ...Vertex Edge System 62 User Manual ...
Страница 69: ...Vertex Edge System 68 User Manual ...
Страница 91: ...Vertex Edge System 90 User Manual ...
Страница 96: ...Vertex Edge System 95 User Manual ...
Страница 98: ...1 Main power ON OFF switch 2 Analyzer power switches Vertex Edge System 97 User Manual ...
Страница 165: ... September 2021 1998 2002_EN_RevA1_VertexEdge_UserManual Vertex Edge System 164 User Manual ...