5
. Internal Memory
5.1. Configuration of internal memory
R200
Isolated Thermocouple Input Module TC4S User's Guide
65
September 2010
Honeywell
Other data monitoring area (use of GET/GETP)
Memory
Address
Description
R/W
Remarks
Dec.
Hex.
60
3CH
CH 0 setting error info (flag)
R
GET
61
3DH
CH 1 setting error info (flag)
62
3EH
CH 2 setting error info (flag)
63
3FH
CH 3 setting error info (flag)
64
40H
CH 0 input change (input change alarm
function data)
R
GET
65
41H
CH 1 input change (input change alarm
function data)
66
42H
CH 2 input change (input change alarm
function data)
67
43H
CH 3 input change (input change alarm
function data)
68
44H
CH 0 disconnection info (code)
R
GET
69
45H
CH 1 disconnection info (code)
70
46H
CH 2 disconnection info (code)
71
47H
CH 3 disconnection info (code)
72
48H
CH 0 cold junction compensating temp (RJC
temperature)
R
GET
73
49H
CH 1 cold junction compensating temp (RJC
temperature)
74
4AH
CH 2 cold junction compensating temp (RJC
temperature)
75
4BH
CH 3 cold junction compensating temp(RJC
temperature)
Содержание TC4S
Страница 114: ...7 Troubleshooting 7 4 Error code 114 Isolated Thermocouple Input Module TC4S User s Guide R200 Honeywell September 2010 ...
Страница 126: ...8 Appendix 8 3 Dimension 126 Isolated Thermocouple Input Module TC4S User s Guide R200 Honeywell September 2010 ...
Страница 127: ...8 Appendix 8 3 Dimension R200 Isolated Thermocouple Input Module TC4S User s Guide 127 September 2010 Honeywell ...
Страница 128: ...Honeywell Process Solutions 1860 W Rose Garden Lane Phoenix AZ 85027 USA ...