![Honeywell TC4S Скачать руководство пользователя страница 122](http://html.mh-extra.com/html/honeywell/tc4s/tc4s_user-manual_151401122.webp)
8
. Appendix
8.2. Thermo electromotive force and compensating cable
122
Isolated Thermocouple Input Module TC4S User's Guide
R200
Honeywell
September 2010
Symbol of
materials
Former
symbols
(cf)
Temperature
Grade
Allowance
K
CA
0
°
C ~ lower
than 1000
°
C
0.4
±
1.5
°
C or
±
0.4% of
temperature measured
0
°
C ~ lower than
1200
°
C
0.75
±
2.5
°
C or
±
0.75% of
temperature measured
-200
°
C~ lower
than 0
°
C
1.5
±
2.5
°
C or
±
1.5% of
temperature measured
E
CRC
0
°
C~ lower than
800
°
C
0.4
±
1.5
°
C or
±
0.4% of
temperature measured
0
°
C~ lower than
800
°
C
0.75
±
2.5
°
C or
±
0.75% of
temperature measured
-200
°
C~ lower
than 0
°
C
1.5
±
2.5
°
C or
±
1.5% of
temperature measured
J
IC
0
°
C~ lower than
750
°
C
0.4
±
1.5
°
C or
±
0.4% of
temperature measured
0
°
C~ lower than
750
°
C
0.75
±
2.5
°
C or
±
0.75% of
temperature measured
T
CC
0
°
C~ lower than
350
°
C
0.4
±
0.5
°
C or
±
0.4% of
temperature measured
0
°
C~ lower than
350
°
C
0.75
±
1
°
C or
±
0.75% of
temperature measured
-200
°
C~ lower
than 0
°
C
1.5
±
1
°
C or
±
1.5% of
temperature measured
ATTENTION
Allowance refers to the allowable maximum limit subtracting the actual
temperature of junction from the converted temperature, based on thermo
electromotive force table of standards. In addition, the allowance will be
greater than
°
C or %.
Содержание TC4S
Страница 114: ...7 Troubleshooting 7 4 Error code 114 Isolated Thermocouple Input Module TC4S User s Guide R200 Honeywell September 2010 ...
Страница 126: ...8 Appendix 8 3 Dimension 126 Isolated Thermocouple Input Module TC4S User s Guide R200 Honeywell September 2010 ...
Страница 127: ...8 Appendix 8 3 Dimension R200 Isolated Thermocouple Input Module TC4S User s Guide 127 September 2010 Honeywell ...
Страница 128: ...Honeywell Process Solutions 1860 W Rose Garden Lane Phoenix AZ 85027 USA ...