![Honeywell SmartLine RM60 Скачать руководство пользователя страница 88](http://html.mh-extra.com/html/honeywell/smartline-rm60/smartline-rm60_handbook_151887088.webp)
8
TECHNICAL DATA
88
SMARTLINE® RM60 (FMCW)
www.honeywell.com/ps
34-VF-25-31 iss.4 GLO December 15 US
8.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website.
Measuring system
Measuring principle
2-wire loop-powered level transmitter; K-band (24...26 GHz) FMCW radar
Application range
Level measurement of powders and granulates
Primary measured value
Δ
f (change in frequency) between the emitted and received signal
Secondary measured value
Distance, level, volume, mass and reflectivity
Design
Construction
The measurement system consists of a measuring sensor (antenna) and a signal
converter which is only available in a compact version
Standard
Antenna purging system for horn antenna (supplied with a
¼
NPTF connection
–
for
horn antenna only)
Options
Integrated LCD display with sun cover (-20..+60
°
C / -4
…
+140
°
F); if the ambient
temperature is not in these limits, the display switches off
2nd current output
PTFE/PP flange plate protection (for Drop antenna without antenna extensions only)
Distance piece (for process temperature: +150...+200
°
C / +300...+390
°
F)
Accessories
Weather protection
Antenna extensions of 105 mm / 4.1
¨
length (max. length for Drop antenna versions:
525 mm / 20.7
¨
)
2
°
slanted PP flange (for all antennas)
Discs (low-pressure flanges) with bolt hole dimensions and positions that agree
with DN80...200 in PN2.5...40 or 3
¨
...8
¨
in 150 lb for devices with the G 1
½
threaded
connection. Max. pressure: 1 barg / 14.5 psig at +20
°
C / +68
°
F.
Max. measuring range
80 m / 260 ft
Depends on the antenna option, dielectric constant of the product and installation
type. Refer also to "Antenna selection".
Min. tank height
0.2 m / 8
¨
Min. dead zone
Antenna extension antenna 0.3 m / 12
¨
Beam angle of antenna
Horn / Sheet metal horn DN80 / 3
¨
: 10
°
Horn / Sheet metal horn DN100 / 4
¨
: 8
°
Sheet metal horn DN150 / 6
¨
: 6
°
Sheet metal horn DN200 / 8
¨
: 4
°
Drop DN80 / 3
¨
: 8
°
Drop DN150 / 6
¨
: 4
°
Display and user interface
Display and user interface
Display and user interface
Display and user interface
Display
LCD display
9 lines, 160 × 160 pixels in 8-step grayscale with 4-button keypad
Interface languages
English, German, French, Italian, Spanish, Portuguese, Japanese, Simplified
Chinese and Russian
Содержание SmartLine RM60
Страница 4: ...4 www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ABOUT THIS DOCUMENT SMARTLINE RM60 FMCW ...
Страница 8: ...CONTENTS 8 www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US SMARTLINE RM60 FMCW ...
Страница 16: ...1 SAFETY INSTRUCTIONS 16 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 34: ...3 INSTALLATION 34 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 40: ...4 ELECTRICAL CONNECTIONS 40 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 76: ...6 OPERATION 76 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 86: ...7 SERVICE 86 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 104: ...8 TECHNICAL DATA 104 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 114: ...9 DESCRIPTION OF HART INTERFACE 114 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 126: ...10 APPENDIX 126 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 127: ...APPENDIX 10 127 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...