![Honeywell SmartLine RM60 Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/honeywell/smartline-rm60/smartline-rm60_handbook_151887041.webp)
START-UP
5
41
SMARTLINE® RM60 (FMCW)
www.honeywell.com/ps
34-VF-25-31 iss.4 GLO December 15 US
5.1 Start-up checklist
Check these points before you energize the device:
•
Are all the wetted components (antenna, flange and gaskets) resistant to the product in the
silo?
•
Does the information on the signal converter nameplate agree with the operating data?
•
Did you correctly install the device on the silo?
•
Do the electrical connections agree with the national electrical codes? Use the applicable
electrical cables with the cable glands.
5.2 Operating concept
You can read measurements and configure the device with:
•
A digital display screen (optional).
•
A connection to a system or PC with PACTware
™
. You can download the Device Type
Manager (DTM) file from the website. It is also supplied on the DVD-ROM delivered with the
device.
•
A connection to a system or PC with AMS
™
. You can download the Device Description (DD)
file from the website. It is also supplied on the DVD-ROM delivered with the device.
•
A connection to a HART
®
Field Communicator. You can download the Device Description (DD)
file from the website. It is also supplied on the DVD-ROM delivered with the device.
DANGER!
Before you energize the device, make sure that the supply voltage and polarity are correct.
DANGER!
Make sure that the device and the installation agrees with the requirements of the Ex certificate
of compliance.
Содержание SmartLine RM60
Страница 4: ...4 www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ABOUT THIS DOCUMENT SMARTLINE RM60 FMCW ...
Страница 8: ...CONTENTS 8 www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US SMARTLINE RM60 FMCW ...
Страница 16: ...1 SAFETY INSTRUCTIONS 16 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 34: ...3 INSTALLATION 34 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 40: ...4 ELECTRICAL CONNECTIONS 40 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 76: ...6 OPERATION 76 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 86: ...7 SERVICE 86 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 104: ...8 TECHNICAL DATA 104 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 114: ...9 DESCRIPTION OF HART INTERFACE 114 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 126: ...10 APPENDIX 126 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Страница 127: ...APPENDIX 10 127 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...