MicroDock II Automatic Test and Calibration Station
59
Troubleshooting
If a problem is encountered, refer to the solutions provided in the following table. If the
problem cannot be resolved, contact your local Distributor or Honeywell Analytics.
Problem
Possible Cause
Solution
The station does not activate.
There is no power connection. Connect the power adapter.
Batteries are not inserted.
Insert the batteries.
The batteries are depleted.
Replace the batteries.
The station is damaged or
defective.
Call for Service.
The detector fails a bump test. Barbed fittings are not connected
correctly with the gasket.
Confirm connection - refer to
Adding Another Docking
Module
.
Detector alarm setpoints
are set higher than the gas
concentration levels of the gas
cylinder.
Connect a gas cylinder that has
a higher gas concentration level
than the alarm setpoints.
Gas connections are not
attached correctly or are
blocked.
Confirm that all gas connections
are attached correctly.
Gas inlets are not setup
correctly.
Refer to
Inlet Setup
and
Installation
.
Gas cylinder is empty.
Use a new gas cylinder.
Detector and/or station are
damaged or defective.
Call for Service.
The oxygen sensor fails a bump
check.
Oxygen alarm setpoints of the
detector are set to the same
concentration level as the gas
cylinder.
Connect a quad-gas cylinder
that has 18% O
2
.
Ambient air (inlet 1) is blocked or
the inlet filter is contaminated.
Remove the blockage or replace
the inlet filter.
Zero gas cylinder is empty.
Use a new zero gas cylinder.
The detector fails a calibration. Gas cylinder is empty.
Use a new gas cylinder.
Barbed fittings are not connected
correctly to the gasket.
Confirm connection - refer to
Adding Another Docking
Module.
Gas connections are not
attached correctly or are
blocked.
Confirm that all gas connections
are attached correctly
Detector and/or station are
damaged or defective.
Call for Service.
Flashing battery icon .
Batter ies are extremely
depleted.
Replace the batteries.
Table 20. Troubleshooting Tips
Содержание MicroDock II
Страница 1: ...MicroDock II Operating Manual MICRODOCK II AUTOMATIC TEST AND CALIBRATION ...
Страница 2: ......
Страница 8: ...MicroDock II Automatic Test and Calibration Station 8 ...
Страница 38: ...MicroDock II Automatic Test and Calibration Station 38 ...
Страница 64: ...MicroDock II Automatic Test and Calibration Station 64 ...
Страница 67: ......