MicroDock II Automatic Test and Calibration Station
18
Inserting the X5
CAUTION
Infrared or intense ambient light (sun or halogen) may interfere with the station/
detector communication.
Figure 6. Inserting the X5 Figure 7. Removing the Diffusion Adapter
NOTe
Table 9 references Figure 6 and Figure 7.
Item Description
1
Docking module lid
2
Release tabs
3
Diffusion adapter
4
Detector bay
5
Connector pins (charging model only)
6
Charge status indicator (charging model only)
7
Connector outlets
8
Diffusion adapter release tab
9
Tool used to push down on diffusion adapter release tab
Table 9. Inserting the X5
Activate the detector and wait until it is in normal operating mode.
1.
Ensure the alligator clip is closed and the ring is resting flat on the detector
2.
to prevent disruptions with the transmission.
Press the two release tabs on the docking module and raise the lid.
3.
Important: If the X5 detector is fitted with a pump, the diffusion adapter
must be removed from the docking module lid.