MicroDock II Automatic Test and Calibration Station
49
If there are no gas conflicts, the station then applies the gas.
If more than one gas is being applied, the station performs a purge between
each gas. Depending upon the gas type, the purge time(s) will vary.
After the bump check is complete, the station displays the results of the
8.
bump check and performs a final purge to clear any remaining gas.
The countdown of seconds remaining displays beside Purge.
Bump Check Results
After the bump check is complete, the station displays the results of the test.
Pass
Fail
H2S –
Gas type not detected (H
2
S is used as an example, this applies to
all gases)
–
Sensor is disabled
NOTe
After a bump check is performed, the X5 deactivates in 5 minutes if no
activity is detected.
Bump Pass
indicates that the bump test has passed. The detector is now ready for use. Press
OK to return to the normal operating screen.
Bump Fail
If a bump test fails,
displays beside the applicable gas (e.g., H
2
S
).
If an audible or visual test fails,
displays beside the applicable test (AUD or VIS).