MicroDock II Automatic Test and Calibration Station
31
To enter a lot number for the corresponding gas cylinder, proceed to
14.
Gas
Cylinder Lot # Field.
Or
Press
exit to return to the user options screen.
Press
15.
exit again to return to the normal operating screen.
Gas Cylinder Lot # Field
Although this field is designed to enter the lot number of the corresponding gas
cylinder, it can be used to enter other data. A maximum of fourteen characters
(letters and/or numbers) can be selected.
Press
16.
to move
down to the lot number field. Initially this field is
blank.
Press
17.
sel. The
changes to * and the cursor displays for the first
value.
Press
16.
or
^ to scroll to the first desired letter/number. Press
sel to
accept the value and automatically move the cursor to the next space.
Repeat step #18 for the remaining values.
17.
The following screen displays the corresponding lot number for the attached
gas cylinder.
Press
20.
exit. The * changes back to the icon.
To setup another inlet, repeat the procedures listed in the
21.
Gas Type
,
Gas
Concentration Level
, and the Gas Cylinder Lot # Field sections.
When all of the entries are completed, press
22.
exit to return to the user
options menu.
Press
23.
exit again to return to the normal operating screen.
Pump Setup
Refer to
Pump Setup
in the
Adding Another Docking Module
section.
Содержание MicroDock II
Страница 1: ...MicroDock II Operating Manual MICRODOCK II AUTOMATIC TEST AND CALIBRATION ...
Страница 2: ......
Страница 8: ...MicroDock II Automatic Test and Calibration Station 8 ...
Страница 38: ...MicroDock II Automatic Test and Calibration Station 38 ...
Страница 64: ...MicroDock II Automatic Test and Calibration Station 64 ...
Страница 67: ......