MicroDock II Automatic Test and Calibration Station
61
Problem
Possible Cause
Solution
No drive letter is created for the
MicroDock II (Windows XP).
Windows XP is not correctly
mapping the drive.
Ensure Windows XP assigned a
drive letter to the MicroDock II.
Solution:
1) Right-click
My Computer
and then click
Manage
.
2 ) U n d e r
C o m p u t e r
Management (Local)
click
Disk Management
3) In the list of drives in the right
pane, right-click the new drive
and then click
Change Drive
Letter and Path(s)
.
4) Click
Change
and in the
drop-down box, select a drive
letter for the new drive (cannot
be assigned to a mapped
network drive).
5) Click
OK
and then click
OK
again.
Problem transferring to the
MMC card during a calibration
or bump test.
Repeat the test.
Ensure the MMC card is
inserted.
Size of MMC is not supported. Use only 128MB
Delkin Multi-
media card (MMC)
M M C i s n o t c o r r e c t l y
formatted.
Refer to
Format the MMC
Card.
Replace with a new MMC
card.
Log file has exceeded maximum
size.
Refer to
Format the MMC
Card.
Replace with a new MMC
card.
IR not working when detector is
inserted into the module.
Check for dirt on the sensor.
Confir m the detector is
activated.
Charge the detector battery if
too low.
Table 20. Troubleshooting Tips
Содержание MicroDock II
Страница 1: ...MicroDock II Operating Manual MICRODOCK II AUTOMATIC TEST AND CALIBRATION ...
Страница 2: ......
Страница 8: ...MicroDock II Automatic Test and Calibration Station 8 ...
Страница 38: ...MicroDock II Automatic Test and Calibration Station 38 ...
Страница 64: ...MicroDock II Automatic Test and Calibration Station 64 ...
Страница 67: ......