5-10
Chadwick-Helmuth A Business Unit of Honeywell International, Inc.
Chapter 5
- Maintenance And Repair
g.
Press
STOP
Key on 8500C/C+.
5.4.1.6 FasTrak
5.4.1.6.1 Test Setup (Ref. Figure 5-6)
NOTE: This test provides a high level of confidence that the FasTrak is operating properly
and is included here for completeness. The associate circuitry in the 8520 and 8500 has
already been tested.
Before doing the FasTrak test, apply the special overlay to the calibrator disk. See Figure 5-6 for
samples that can be copied or recreated.
NOTE: This test should be performed in a well-lit room.
a.
Place calibrator on its back on a flat surface.
b.
Place FasTrak on its side with the pin connector facing to the right. The window should be
facing the calibrator approximately six inches away.
c.
Set
CAM RATE (RPM)
switch to
900
.
d.
Set
MOTOR
switch to
ON
.
e.
Adjust pot next to
900 CAM RATE
switch position until 2nd ring from outside on rotor
disk appears frozen in place under 60 Hz light.
3
f.
Press
BALANCE
key on 8500C/C+.
g.
Press
SETUP
key.
h.
Select
PICKUP
field with arrow keys.
i.
Press
CHANGE
soft key to select
PICKUP B
.
j.
Press
STATUS
key.
k.
After a few seconds, the RPM on the screen should read approximately 900 RPM.
NOTE: If you cannot adjust the distance or alignment to obtain 900 RPM, there may be a
problem with the FasTrak. Return it to Chadwick-Helmuth for adjustment.
l.
With a white index card or piece of paper, cover half of the rotating disk. Place the card
close to, but not touching, the disk, so as to cover 1/2 Mag Pickup sensor and one vibration
sensor.
m.
The RPM should drop by approximately half, remove the card.
Night Mode Test
a.
To simulate night operation, turn off all lights in room, or enclose the test setup within a
darkened enclosure, such as a large box. The RPM should drop to zero when the area is
dark enough.
b.
Switch Tracker switch to Night mode. (in the cable)
c.
After a few seconds, the RPM should appear on the display
3
Attention, users outside U.S.: The 11A Calibrator can also be calibrated under 50Hz light. For 1714
rpm and 923 rpm cam rates, inner two rings on rotor disk are used. Remember to enter half these rates
in 8500C/C+ setup screen, when called for in the following steps.
Содержание Chadwick-Helmuth 8500C
Страница 12: ...xii Chadwick Helmuth A Business Unit of Honeywell International Inc List of Tables ...
Страница 17: ...System Physical Description 8500C 8500C System Maintenance Manual 1 5 Figure 1 3 System Cables ...
Страница 44: ...4 14 Chadwick Helmuth A Business Unit of Honeywell International Inc Chapter 4 System Theory of Operation ...
Страница 97: ...Disassembly 8500C 8500C System Maintenance Manual 5 53 Figure 5 17 Balancer Analyzer Assembly Parts Breakdown ...
Страница 103: ...Disassembly 8500C 8500C System Maintenance Manual 5 59 Figure 5 19 8500C C Harness Assembly Sheet 1 of 2 ...
Страница 105: ...Disassembly 8500C 8500C System Maintenance Manual 5 61 Figure 5 19 8500C C Harness Assembly Sheet 2 of 2 ...
Страница 109: ...Disassembly 8500C 8500C System Maintenance Manual 5 65 Figure 5 21 8520C Harness Assemblies Sheet 1 of 2 ...
Страница 111: ...Disassembly 8500C 8500C System Maintenance Manual 5 67 Figure 5 21 8520C Harness Assemblies Sheet 2 of 2 ...
Страница 122: ...5 78 Chadwick Helmuth A Business Unit of Honeywell International Inc Chapter 5 Maintenance And Repair ...
Страница 154: ...A 32 Chadwick Helmuth A Business Unit of Honeywell International Inc Appendix A 8520C 36 Signal Selector ...
Страница 179: ...Maintenance and Repair 8500C 8500C System Maintenance Manual D 21 Figure D 12 Wiring Diagram 8520CS Sheet 1 of 2 ...
Страница 181: ...Maintenance and Repair 8500C 8500C System Maintenance Manual D 23 Figure D 12 Wiring Diagram 8520CS Sheet 2 of 2 ...