ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
9.9
9
T
ROUBLESHOOTING
5015 - Duplicate
Slide ID
Multiple slides were labeled
with the same ID number in the
batch. The vial with the dupli-
cate ID will not be processed.
In Small Batch Mode, multiple
slides were labeled with the
same ID number.
In AutoLoader mode, the ID for
a slide was generated but not
etched because the slide ID
would have the same ID num-
ber as another slide ID.
The vial ID and/or slide ID bar-
code configuration and/or the
slide label design is/are not set
up to correctly identify the sec-
tion of the vial label that is the
accession ID and etch it on the
slide.
Check the sample IDs and con-
firm that they are duplicates. A
slide was only made from the
first vial.
The patient information must
be checked and reconciled for
both vials.
In Small Batch mode, relabel
the second slide and repro-
cess.
In AutoLoader mode, check
and correct the Label Design.
Correct the slide ID barcode
configuration on the instru-
ment. Refer to “Configure Slide
ID” on page 6.51.
5016 - Sample
Error During
Processing
Positive tank error or other
error during cell transfer.
This message is only a notifica-
tion; the slide is processed and
may be adequate.
The filter might be punctured
or defective.
The system has a pressure
leak.
Gyn slides -
If the slide is satisfactory for
screening purposes, no further
action is necessary.
If the slide is inadequate, follow
laboratory procedure for
reporting unsatisfactory speci-
mens.
Non-gyn slides -
If there is additional sample
material available, make
another slide with more cells if
possible.
If the error persists, contact
Technical Support.
Table 9.1 Sample Processing Errors
Error
Description
Possible Cause
Corrective Action
Содержание ThinPrep 5000
Страница 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Страница 5: ...This page intentionally left blank...
Страница 6: ...Instructions For Use Instructions For Use...
Страница 43: ...Table of Contents Table of Contents...
Страница 48: ...1 Introduction 1 Introduction...
Страница 73: ...2 Installation 2 Installation...
Страница 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Страница 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Страница 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Страница 134: ...6 User Interface 6 User Interface...
Страница 216: ...USER INTERFACE 6 82 ThinPrep 5000 Processor with AutoLoader Operator s Manual 6 This page intentionally left blank...
Страница 217: ...7 Operating Instructions 7 Operating Instructions...
Страница 256: ...8 Maintenance 8 Maintenance...
Страница 265: ...ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 9 8 MAINTENANCE Figure 8 8 Slide Waste Bin...
Страница 282: ...MAINTENANCE 8 26 ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 This page intentionally left blank...
Страница 283: ...9 Troubleshooting 9 Troubleshooting...
Страница 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Страница 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Страница 316: ...Service Information Service Information...
Страница 319: ...Ordering Information Ordering Information...
Страница 326: ...Index Index...
Страница 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...