5.12
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
CYTOLYT SOLUTION WASH
Addition of CytoLyt
™
Solution to cell pellets is required to wash the sample. A
CytoLyt Solution
Wash
performs the following functions while preserving cellular morphology:
•
Lyse red blood cells
•
Dissolve mucus
•
Reduce protein precipitation
A
CytoLyt Solution Wash
consists of the following process:
•
Adding 30 mL of CytoLyt Solution to a cell pellet
•
Mucoid Specimens Only: Mechanical agitation
•
Concentration by centrifugation — 600g x 10 minutes
•
Pouring off the supernatant and vortexing to resuspend the cell pellet
One
CytoLyt Solution Wash
is usually adequate to clean most non-gyn samples. For particularly
bloody or mucoid specimens, additional
CytoLyt Solution Washes
may be necessary.
When a sample is collected in CytoLyt Solution at a ratio less than 30 parts CytoLyt Solution to 1 part
sample, this is considered a
Collection Step
and not a
Wash Step
. For example, if one collects 15 mL of a
sample and adds 30 mL of CytoLyt Solution to this sample, then the CytoLyt Solution: sample ratio is
only 2 to 1 and this is considered a sample collection step and still requires a
CytoLyt Solution Wash
.
For more information on CytoLyt Solution, refer to Chapter 3, PreservCyt™ & CytoLyt™ Solutions.
MUCOID
ONLY
Содержание ThinPrep 5000
Страница 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Страница 5: ...This page intentionally left blank...
Страница 6: ...Instructions For Use Instructions For Use...
Страница 43: ...Table of Contents Table of Contents...
Страница 48: ...1 Introduction 1 Introduction...
Страница 73: ...2 Installation 2 Installation...
Страница 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Страница 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Страница 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Страница 134: ...6 User Interface 6 User Interface...
Страница 216: ...USER INTERFACE 6 82 ThinPrep 5000 Processor with AutoLoader Operator s Manual 6 This page intentionally left blank...
Страница 217: ...7 Operating Instructions 7 Operating Instructions...
Страница 256: ...8 Maintenance 8 Maintenance...
Страница 265: ...ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 9 8 MAINTENANCE Figure 8 8 Slide Waste Bin...
Страница 282: ...MAINTENANCE 8 26 ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 This page intentionally left blank...
Страница 283: ...9 Troubleshooting 9 Troubleshooting...
Страница 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Страница 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Страница 316: ...Service Information Service Information...
Страница 319: ...Ordering Information Ordering Information...
Страница 326: ...Index Index...
Страница 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...