ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
7
CAUTION:
For best slide preparation results, use the correct slide and filter type for the sample
type that is processed.
Table 7.1: Sample/Filter/Slide Configurations
When sample vials, filters and slides are loaded into the system, the operator must identify what
type they are via the user interface.
Load Slide Cassettes with Microscope Slides
A slide cassette holds approximately 100 slides; the equivalent of an entire box of packaged ThinPrep
microscope slides. The slide cassettes have a highly polished inner surface, to help the slide handler
pick slides without binding. Before loading a cassette with microscope slides, gently wipe down the
inside surface with a soft cloth to remove any glass dust from previous loads.
The most important part of loading slides into the cassette is to get them oriented correctly, so that
the frosted area of the slide receives the laser etching. The frosted area must be facing up and toward
the rear of the cassette. As a user aid, the interior of the cassette has graphic instruction for loading
slides etched onto its surface.
ThinPrep
Th Imaging
UroCyte
PreservCyt Sam-
ple
Gynecologic
Non-gynecologic
Gynecologic
Urine for use with Vysis
UroVysion molecular test-
ing
Filter
Clear
Blue
Clear
Yellow
Slide
Cell spot arc
Cell spot arc or arc-less
Cell spot arc with fidu-
cial marks
Cell spot circle
Содержание ThinPrep 5000
Страница 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Страница 5: ...This page intentionally left blank...
Страница 6: ...Instructions For Use Instructions For Use...
Страница 43: ...Table of Contents Table of Contents...
Страница 48: ...1 Introduction 1 Introduction...
Страница 73: ...2 Installation 2 Installation...
Страница 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Страница 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Страница 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Страница 134: ...6 User Interface 6 User Interface...
Страница 216: ...USER INTERFACE 6 82 ThinPrep 5000 Processor with AutoLoader Operator s Manual 6 This page intentionally left blank...
Страница 217: ...7 Operating Instructions 7 Operating Instructions...
Страница 256: ...8 Maintenance 8 Maintenance...
Страница 265: ...ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 9 8 MAINTENANCE Figure 8 8 Slide Waste Bin...
Страница 282: ...MAINTENANCE 8 26 ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 This page intentionally left blank...
Страница 283: ...9 Troubleshooting 9 Troubleshooting...
Страница 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Страница 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Страница 316: ...Service Information Service Information...
Страница 319: ...Ordering Information Ordering Information...
Страница 326: ...Index Index...
Страница 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...