N
ON
-G
YNECOLOGIC
S
AMPLE
P
REPARATION
5.26
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
Techniques Used in Troubleshooting
Diluting the Sample 20 to 1
To dilute a sample suspended in PreservCyt Solution, add 1 mL of the sample that is suspended in
PreservCyt Solution to a new PreservCyt Solution vial (20 mL). This is most accurately done with a
calibrated pipette.
You may also simply count drops from an uncalibrated plastic pipette if you know how many drops
correspond to 1 mL. To calculate this, count out drops of PreservCyt Solution into a container of
known volume. When the known volume is reached, divide the number of drops by the volume (in
mL) to get the number of drops that corresponds to 1 mL. Use PreservCyt Solution rather than any
other liquid so the drop size will be consistent with the sample drops.
Glacial Acetic Acid Wash for Blood and Non-Cellular Debris
If a sample is found to be bloody during microscopic review, it can be further washed using a
solution of 9 parts CytoLyt Solution and 1 part glacial acetic acid. This should only be done after the
sample has been in PreservCyt Solution. Do not use directly with fresh specimens; nuclear
morphology may not be adequately preserved.
6. Add specimen to PreservCyt™ Solution
vial (page 5.10).
7. Run on ThinPrep
™
5000 processor with
AutoLoader using Sequence Non-Gyn. Fix,
stain, and evaluate.
8. If the new slide is sparse, call Hologic
Technical Support (page 12.1).
Problem
Procedure
Содержание ThinPrep 5000
Страница 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Страница 5: ...This page intentionally left blank...
Страница 6: ...Instructions For Use Instructions For Use...
Страница 43: ...Table of Contents Table of Contents...
Страница 48: ...1 Introduction 1 Introduction...
Страница 73: ...2 Installation 2 Installation...
Страница 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Страница 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Страница 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Страница 134: ...6 User Interface 6 User Interface...
Страница 216: ...USER INTERFACE 6 82 ThinPrep 5000 Processor with AutoLoader Operator s Manual 6 This page intentionally left blank...
Страница 217: ...7 Operating Instructions 7 Operating Instructions...
Страница 256: ...8 Maintenance 8 Maintenance...
Страница 265: ...ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 9 8 MAINTENANCE Figure 8 8 Slide Waste Bin...
Страница 282: ...MAINTENANCE 8 26 ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 This page intentionally left blank...
Страница 283: ...9 Troubleshooting 9 Troubleshooting...
Страница 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Страница 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Страница 316: ...Service Information Service Information...
Страница 319: ...Ordering Information Ordering Information...
Страница 326: ...Index Index...
Страница 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...