ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
9.7
9
T
ROUBLESHOOTING
5012 - Vial/Slide ID
Mismatch (small
batch mode only)
The vial and slide IDs were
both successfully read, but did
not match. The sample was not
processed and no slide was
made.
Slides placed in wrong carou-
sel slot.
Incorrect labeling of slides or
vials.
The slide ID barcode configura-
tion is not set up to correctly
identify the section of the vial
label that is the accession ID
for the sample.
Examine the vial and slide IDs
to confirm they do not match.
See if the slide has been filed
in the wrong slot on the carou-
sel. (Look at subsequent IDs, in
case the mistake was perpetu-
ated within the carousel.)
Reconcile the patient informa-
tion with the correct ID. Rela-
bel, if necessary.
Correct the slide ID barcode
configuration on the instru-
ment. Refer to “Configure slide
ID for pre-labeled slides” on
page 6.59
5013 - End of Vial
or Filter Not
Present in Multiple
Slides Per Vial
(small batch mode
only)
The entire sample was con-
sumed during the advanced
process sequence ‘Multiple
Slides Per Vial’. This error only
occurs during Multiple Slides
Per Vial mode, which does not
check for fluid level or dilute
sample. The slide was pro-
cessed, but should be checked
for adequacy.
Filter not present.
All fluid in the vial was con-
sumed.
Pneumatic system failure.
Make sure a filter is present.
If Multiple slides per vial mode
is being used, there is not
enough sample to process the
desired number of slides.
Examine the vial to see if it is
empty.
Table 9.1 Sample Processing Errors
Error
Description
Possible Cause
Corrective Action
Содержание ThinPrep 5000
Страница 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Страница 5: ...This page intentionally left blank...
Страница 6: ...Instructions For Use Instructions For Use...
Страница 43: ...Table of Contents Table of Contents...
Страница 48: ...1 Introduction 1 Introduction...
Страница 73: ...2 Installation 2 Installation...
Страница 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Страница 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Страница 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Страница 134: ...6 User Interface 6 User Interface...
Страница 216: ...USER INTERFACE 6 82 ThinPrep 5000 Processor with AutoLoader Operator s Manual 6 This page intentionally left blank...
Страница 217: ...7 Operating Instructions 7 Operating Instructions...
Страница 256: ...8 Maintenance 8 Maintenance...
Страница 265: ...ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 9 8 MAINTENANCE Figure 8 8 Slide Waste Bin...
Страница 282: ...MAINTENANCE 8 26 ThinPrep 5000 Processor with AutoLoader Operator s Manual 8 This page intentionally left blank...
Страница 283: ...9 Troubleshooting 9 Troubleshooting...
Страница 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Страница 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Страница 316: ...Service Information Service Information...
Страница 319: ...Ordering Information Ordering Information...
Страница 326: ...Index Index...
Страница 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...