System User Manual
Section 4 —
Software Functions — CHANGE SETUP — Rapid fFN 10Q QCette® SETUP
The USER ID must be entered to proceed to the next step. Press
ENTER to accept the ID. This fi eld will accept 15 alpha or numeric
characters.
The QCette SN (serial number) must be entered to proceed to the
next step. The serial number is printed on the QCette plastic housing.
The software requires that the serial number is entered in the correct
format: six numeric characters (e.g., 004640). Enter all leading zeros.
This message will be displayed if a cassette is present in the analyzer
prior to reaching the next screen. Remove cassette and press ENTER.
The analyzer then prompts the user to insert the QCette and press
ENTER. A two-minute timer starts during which time this message
fl ashes and the analyzer beeps. Insert the QCette and press ENTER.
QCette SETUP
USER ID
>JOHN SMITH
ENTER-CONFIRM
QCette SETUP
ENTER QCette SN
>004640
ENTER-CONFIRM
REMOVE CASSETTE
PRESS ENTER
TO CONTINUE
INSERT QCette
ENTER-CONTINUE