4–29
Rapid fFN 10Q
System User Manual
Section 4 —
Software Functions — CHANGE SETUP — Rapid fFN 10Q QCette® SETUP
CHANGE SETUP — Rapid fFN 10Q QCette
®
SETUP
Option 6 on the Main Menu screen allows the user to change the
Date/Time, Autoprint, or to perform QCette SETUP from the SETUP
MENU. This option also allows Hologic Technical Support to update
the software version.
The QCette SETUP initializes the QCette for use in evaluating the
performance of the analyzer. During the initialization process, the
performance criteria of the analyzer are established. QCette SETUP
must be performed PRIOR to running the QCette as a quality
control device.
Select option 3 on the SETUP MENU to begin QCette SETUP.
fFN MAIN MENU
4-ACCESS DATA
5-VIEW SETUP
6-CHANGE SETUP
SETUP MENU
1-DATE/TIME
2-AUTOPRINT
3-QCette SETUP