7–4
Rapid fFN 10Q
System User Manual
Section 7 — Troubleshooting
ITEM
PROBLEM
SOURCE
SOLUTION
6
Printer output in
unusual font.
Printer
Disconnect the printer power cord from the electrical outlet.
Turn off the analyzer.
Reconnect the printer power cord to the electrical outlet.
Turn on the analyzer.
7
Analyzer
turned off after
calibration, or
power failure
occurred after
calibration.
Power
The calibration remains in memory. Reset calibration only if
prompted by analyzer.
8
Cassette or
QCette cannot
be removed from
the analyzer.
(
Note:
do not
try to force or
pry it out of the
analyzer.)
Analyzer
Turn analyzer off and back on to reinitialize the system.
If the cassette/QCette can be easily removed, proceed with
testing.
If the cassette/QCette cannot be easily removed, call
Technical Support.