7–7
Rapid fFN 10Q
System User Manual
Section 7 — Troubleshooting
ITEM
PROBLEM
SOURCE
SOLUTION
11
Liquid control
result is outside
the range of
acceptable
results.
Liquid Control
Verify that the Rapid fFN 10Q Cassettes have not expired
and have been stored properly.
Ensure the cal code programmed in the instrument matches
the cal code on the cassette box.
Verify the control has not expired, and is neither cloudy nor
discolored.
If contamination of the controls is suspected, open a new
box of controls.
Review the Rapid fFN Control Kit directional insert to ensure
the correct procedure was followed. Repeat the test on
a new cassette. If control result is acceptable, no further
troubleshooting is required.
If the control result is outside the range of acceptable results
a second time, call Technical Support.