4–3
Rapid fFN 10Q
System User Manual
Section 4 —
Software Functions — SET CALIBRATION
SET CALIBRATION
Option 2 on the Main Menu screen allows the user to set the calibration on
the analyzer. Follow the analyzer prompts. Calibration must be set when
changing cassette lots.
Note:
If the calibration has not been set, menu option 2 will fl ash.
Calibration must be set before the analyzer can be used for testing.
The USER ID must be entered to proceed to the next step. Press ENTER to
accept the ID. This fi eld will accept 15 alpha or numeric characters.
SET CALIBRATION
USER ID
>JOHN SMITH
ENTER-CONFIRM
fFN MAIN MENU
1-TEST PATIENT
2-SET CALIBRATION
3-DAILY QC